23
7-5-3 Zero calibration
4. With nothing placed on the load cell, press the
1
↵
key.
1
↵
Performs zero calibration and goes to step 5.
ESC
Cancels the operation and returns to the
fCal
display.
7-5-4 Span calibration
5. Place a weight on the load cell.
Use the following keys to enter the value of the actual weight used.
Wait for the value to stabilize and press the
1
↵
key.
1
>
Selects the digit to set.
1
∧
Increases the value.
1
∨
Decreases the value.
1
↵
Performs span calibration and goes to step 6.
ESC
Cancels the operation and returns to the
fCal
display.
6. After the span value is saved, the value is displayed for two seconds.
When the full calibration is completed,
end
is displayed. Press
the
1
↵
key to return to the
fCal
display.
Press the ESC key to return to the measurement mode.
Span calibration
10000
fCal
5pan
------
fCal
5pan
1.8020
fCal
5pan
end
fCal
fCal
Cal0
fCal
------
fCal
Cal0