5-1-6 DAV output terminals (D/A voltage output) (10) - (12)
(10) FG terminal
Frame ground terminal for the DAV output. Connects the DAV output cable shield here.
(11) DAV output terminal
DAV output is –10 V to +10 V. Scaling is available by the function setting.
(12) DAV output ground terminal
5-1-7 Analog amplifier output terminals (13) - (15)
(13) FG terminal
Frame ground terminal for the analog amplifier output. Connects the analog amplifier output
cable shield here.
(14) Analog amplifier output terminal
Analog amplifier output is –10 V to +10 V at ±3.2 mV/V. Amplifies the sensor inputs in an analog
manner and outputs the amplified signal.
(15) Analog amplifier output signal ground terminal
Loading the analog output
Connect a non-inductive load of 5 k
Ω
or more to the analog output terminals.
DAV and analog amplifier output
The sensor’s analog signals are amplified 625 times and the analog amplifier
output (AAO) outputs the amplified signal directly.
Therefore, scaling is not
available.
The DAV output is the measured value in digital format, converted to an analog
output voltage. Scaling is available. A 2000 times-per-second output is available.