42
10-5 2D Comparator Mode (Additional Explanation)
While the conventional comparison method detects and judges the maximum or minimum value only
during measurement, the 2D comparator mode uses contact inputs from the rear panel terminals or
elapsed time to make a comparison at various levels.
This type of comparison is useful for an operation such that the pressure increases rapidly at the
pressurization starting time and changes in pressure occur during the pressurization process.
Using this mode, an erroneous situation such as the fall over of a rivet which receives the pressure
or assembling error due to the wrong mounting holes can be detected accurately.
Time
シリンダー
圧入対象
0
100 200 300
400
500
600
700
800
900 1000 1100 1200
0
100
200
300
400 500 600 700
3
1
5
2
4
Cylinder
Object that receives pressure
W
eight (Force)