![A&D AD-4532B Instruction Manual Download Page 10](http://html1.mh-extra.com/html/aandd/ad-4532b/ad-4532b_instruction-manual_3079816010.webp)
8
4. FRONT PANEL
Main display
Displays a measured value or set value.
To set the decimal point position, use function mode
f-00
.
Sub-displays
Displays an upper limit value (left side) or lower limit value (right
side), or displays the set value.
Status indicators
HI
: Turns on when the measured value exceeds the upper
limit value (HI).
OK
: Turns on when the measured value is as shown below.
Lower limit value
≤
measured value
≤
upper limit value
LO
: Turns on when the measured value does not reach the
lower limit value (LO).
NG
: Turns on when the comparison result of the 2D
comparator mode is no good.
LATCH : Turns on when the LATCH function is being performed.
PEAK : Turns on when the hold function is started.
HOLD : Turns on when the value is being held.
5-level comparator
indicators
Displays each level and the result when the 2D comparator mode is
in operation.
Stabilization indicator
Turns on when the measured value is stable.
Zero indicator
Turns on when the measured value is at the zero point.
Status indicators
5-level comparator indicators
Stabilization indicator
Zero indicator
Sub-display (left)
Main display
Sub-display (right)
Keys
Standby indicator