7-3-3 Zero calibration
4. With nothing placed on the load cell, press the
1
↵
key.
1
↵
Performs zero calibration and goes to step 5.
ESC
Cancels the operation and returns to the
d-5p
display.
7-3-4 Digital span calibration
5. Use the following keys to enter the sensor’s rated output voltage.
1
>
Selects the digit to set.
1
∧
Increases the value.
1
∨
Decreases the value.
1
↵
Saves the rated output voltage and goes to step 6.
ESC
Cancels the operation and returns to the
d-5p
display.
6.
end
is displayed. Press the
1
↵
key to return to the
d-5p
display.
Press the
1
>
key to go to the calibration mode.
Press the ESC key to return to the measurement mode.
Digital span calibration
------
d-5p
d5pn
32000
d-5p
d5pn
end
d-5p
d-5p
Cal0
d-5p