25
8-2 Description of the Function Items
Item and Parameter
Description
0
No decimal point
e.g.:
123456
1
1 decimal place
e.g.:
12345.6
2
2 decimal places
e.g.:
1234.56
3
3 decimal places
e.g.:
123.456
Decimal
point
f-00
Decimal point position
4
4 decimal places
e.g.:
12.3456
0
3 Hz
1
10 Hz
2
30 Hz
3
100 Hz
4
300 Hz
f-01
Input filter
5
1 kHz
Selects a pass band for the
input analog low pass filter.
f-02
Moving average filter
0
|
254
0
:
No moving average filter
1
to
254
: Number of moving average
0
:
Factory setting
Filter
f-03
Digital filter
0
|
220
0
:
No digital filter
14
to
220
: Specified cutoff frequency (-3dB) selected from
14, 20, 28, 40, 56, 80, 110, 160, 220 Hz.
14
:
Factory setting
0
5 V
Excitation
voltage
f-04
Excitation voltage
1
2.5 V
Selects the voltage supplied
to the sensor.
1
1 time/second
2
2 times/second
4
4 times/second
8
8 times/second
16
1
16 times/second
Display
f-05
Display update rate
32
1
32 times/second
Display rewriting speed of the
measured value.
Zero
adjustment
f-06
Zero adjustment range
0
|
100
Expressed in percent of the rated capacity with the
calibration zero point as the center
30
:
Factory setting
f-07
Stability detection time
0
|
99
0
:
No stability detection
1
to
99
: 0.1 second to 9.9 seconds
10
:
Factory setting
Stability
f-08
Stability detection band
0
|
99
0
:
No stability detection
1
to
99
: 1 digit to 99 digits
5
:
Factory setting
f-09
Zero tracking time
0
|
99
0
:
No zero tracking
1
to
99
: 0.1 second to 9.9 seconds
f-10
Zero tracking band
0
|
99
0
:
No zero tracking
1
to
99
: 1 digit to 99 digits
0
Rated ca8 digits
Zero
racking
f-11
Overflow
1
AD conversion overflow
: Factory settings.