![A&D AD-4532B Instruction Manual Download Page 49](http://html1.mh-extra.com/html/aandd/ad-4532b/ad-4532b_instruction-manual_3079816049.webp)
47
11-2 Digital to Analog Voltage Output (DAV)
Processes the measured values from the sensors according to the values set in function modes
f-22
to
f-24
, and outputs the processed signal as a voltage after D/A conversion. The output
voltage range is from –10 V to +10 V. Scaling is available in function modes
f-22
and
f-23
.
11-2-1 Specifications
Rated output range
–10 V to +10 V (Non-inductive load of 5 k
Ω
or more)
Maximum output range
–11.0 V to +11.0 V
D/A conversion speed
2000 times per second
Output resolution
1/10000
Zero drift
±1 mV/°C typ.
Gain drift
±100ppm/°C typ.
Linearity
±0.05%F.S. (Drift in the sensor input unit not included)
Load resistance
5 k
Ω
or more
11-2-2 Fine adjustment of zero and span
Note on using this function
Use this mode only when adequate accuracy can not be obtained by the regular
scaling set in function modes
f-22
and
f-23
. This mode uses the internal
correction factor and requires a high-precision digital multimeter. If a low-precision
multimeter is used, it may adversely affect the indicator performance and the output
voltage accuracy.
1. In the measurement mode, press the FNC key for three seconds or
more to enter the function selection mode.
2. Press the > key several times to display
dav
. Press the
↵
key to enter the DAV adjustment mode.
FNC
3 seconds
4532
1000 400
fnc
dav