Guidance and manufacturer’s declaration – electromagnetic emission
The Elipar DeepCure-L is intended for use in the electromagnetic environment specified below.
The customer or the user of the Elipar DeepCure-L should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions The Elipar DeepCure-L uses RF energy only for its internal function.
Group 1 Therefor, its RF emissions are very low and are not likely to cause
CISPR 11 any interference in nearby electronic equipment.
RF emissions
Class B
CISPR 11
Harmonic emissions The Elipar DeepCure-L is suitable for use in all establishments,
Class A including domestic establishments and those directly connected
IEC 61000-3-2 to the public low-voltage power supply network that supplies
Voltage fluctuations /
buildings used for domestic purposes.
flicker emissions
Complies
IEC 61000-3-3