Pannoramic SCAN 150 1.17 User’s Guide
6 Technical Data
Light sources
Halogen lamp (transmitted light operation)
HAL 12V, 5 W
Average service life of halogen lamp
4000h
Lamp module for HXP 120 (fluorescent option)
120 W
Mercury short arc reflector lamp average service life
2000h
Lumencor light source
SPECTRA 6 Light Engine
Estimated service life of the Lumencor light source
75 000 h
This User’s Guide describes Lumencor Spectra 6 Light Engine and HXP 120 as light sources. The X-Cite
120 illumination unit may also be used as a substitute light source.
Slide identification
Slides are identified by a VRmagic VRmC-8+ PRO preview camera. Label area is saved as image and
barcode is interpreted as a character string, with capability for interpretation of the following types of
1D and 2D barcodes:
•
Interleaved 2 of 5
•
Code39 (ASCII encoding)
•
Code93 (ASCII encoding)
•
Code128 (UCC/EAN128) (ASCII encoding)
•
EAN8 (Numeric encoding)
•
EAN13 (Numeric encoding)
•
UPC-A (Numeric encoding)
•
DataMatrix (Numeric encoding, Alpha encoding, AlphanumericPunc encoding,
AlphaNumeric encoding, ASCII encoding, IS08 encoding)
•
Maxicode (Encoding mode 2, 3, 4, 5, 6)
•
MicroPDF417 (Standard encoding type)
•
PDF417 (Standard encoding type)
•
QR (QR code Model 1, 2 encoding)
2014/ April 3/
3DHISTECH
Ltd.
128
(
131
)