Pannoramic SCAN 150 1.18 User’s Guide
5 Troubleshooting and Maintenance
Caution!
Incorrect handling may considerably reduce the lamp
module’s useful
life. Hold it by the ceramic part in all cases! Never touch the lamp’s
glass surface or the inside surface of the reflector. Skin oil is harmful to
these parts.
7. Tighten pressure bolt again with the lever (
Figure 79 – 1
).
8. Re-establish plug-in connection.
9. Screw the lateral cover-plate down to the housing.
10. Connect power plug.
11. Set hour meter (time counter) to zero.
5.3.4 Removing a slide
If a slide got jammed or has suffered breakage during the process of digitization, stop the process
and, if possible, move the slide holder to a position where the slide can be easily removed by clicking
the
Emergency Manual Slide Removal Position
).
Warning!
Crush hazard! Before you remove a jammed slide, cancel
the running
digitization process, terminate the software session and turn the power
off.
Caution!
Any manipulation performed on a mechanical component must be
followed by a complete device initialization routine (software restart).
April 3, 2014 - Rev. 1
3DHISTECH Ltd.
119(131)