Pannoramic SCAN 150 1.18 User’s Guide
5 Troubleshooting and Maintenance
4. Remove magazine manually (see section
5.3.6 Removing the Magazine from the
). If a jammed magazine cannot be removed, contact service
personnel.
5. Reattach protective covers.
6. Reconnect 24V power supply and control computer to line power supply.
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
Illumination error! Probably the lamp has burned out. Please turn off the scanner and check the
lamp!
•
Halogen lamp is defective
•
Replace the halogen lamp (see section
).
Image is black
•
Halogen lamp is defective
•
Replace the halogen lamp (see section
).
Virtual slide is improperly focused almost across its entire surface
•
Focus range enabled
•
Disable
Limiting focus range
.
Poor image quality
•
Imaging error (for example, chromatic aberrations)
•
Contact 3DHISTECH Service for support.
Warning!
Do not modify the camera driver settings, as the default factory
settings give the most accurate images.