Pannoramic SCAN 150 1.18 User’s Guide
5 Troubleshooting and Maintenance
2. Pull protective cover halves apart or remove the right part completely (see section
Removing the protective covers
).
3. Select the desired filter in the corresponding software window (see section
Microscope Settings / Filter Settings
). The filter will be positioned in such a way that the
user can access it without having to turn the filter wheel manually.
4. Enter the respective parameters in the input fields.
Warning!
Crush hazard! Rotating the filter turret motorically during installation
or removal may lead to injury.
5. Hold the filter module
(3)
by its handling pads (one on the right, one on the left (lateral)
side of the module) and insert it in forward tilted position from above into the upper spring-
loaded clamps
(1)
at the filter wheel.
6. Press the filter module down, until it safely connects to the bottom spring-loaded clamps
(2)
of the filter wheel. Mounting position of the filters: excitation filter
(5)
; emission filter
(4)
.
7. To remove the filter module, first slightly tilt it out of its upper spring-loaded fixtures, then
pull it out of the bottom spring-loaded components and take it off.
8. Place the protective covers back again (see
5.3.1 Removing the protective covers
).
9. If an HXP 120 illumination unit is used, turn it on again (consider the cooling time possibly
required).
April 3, 2014 - Rev. 1
3DHISTECH Ltd.
122(131)