Operation
6
Cooling Engine Self Test
During SETUP and then every four hours, during run, the System will
dynamically test its cooling engine. This test will take a few minutes.
Cooling Engine Self Test - The Pump Will Stop
Each hour the System tests its cooling engine dynamically during RUN. These
tests are brief. The system will stop the pump during these tests to remove the
heat load from the patient during the test.
THE PUMP WILL STOP DURING THIS TEST - THIS IS NORMAL.
During these tests, the peristaltic pump is stopped (to remove the patient’s heat
load from the equation). The tests that are conducted depend upon the state of
the System at the time of the test. The test(s) done is(are) as follows:
Coolant Temperature
Heating Test
Cooling Test
<10 °C
10 °C - 38 °C
> 38 °C
= will be performed
= will not be performed
You will not see any screen messages unless there is an abnormal result in
which case the appropriate alarm state will be called.
Sensor Checks
Figure 4-1. System Setup Screen
For the system to enter the next state, STANDBY, the following sensors must be
checked and found normal.
600479-001 Rev 2 37/147
Содержание thermogard xp
Страница 1: ...Operation Manual 600479 001 Rev 2 1 147...
Страница 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Страница 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Страница 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Страница 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Страница 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Страница 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Страница 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Страница 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Страница 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Страница 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Страница 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...