WARNING! Change to Initial Specification of System.
In the initial release of the System there were no settable patient temperature
alarms. There were fixed high and low patient temperature alarms. These
fixed alarms have been made programmable.
The patient temperature alarms MAY NOT be deactived. They may be muted
for a 2 minute period.
NOTE: Patient Temperature Alarms in STANDBY
If the System is in STANDBY and the T1 temperature probe is connected to
the System but NOT inserted into the patient you will experience “Lo” patient
temperature alarms if the ambient temperature of the exposed probe is below
the alarm limit. To avoid the annoyance of the alarm under these conditions,
simply unplug the T1 temperature probe from the System until you are ready to
connect to the patient. You may not enter RUN without the probe connected
to the System.
Bath Pre-set
The menu allows the user to select from either: Pre-Cool, Pre-Warm or None.
This selection operates only in STANDBY and is cancelled when the system
enters RUN.
Figure 4-13. Bath Pre-Set Menu
600479-001 Rev 2 48/147
Содержание thermogard xp
Страница 1: ...Operation Manual 600479 001 Rev 2 1 147...
Страница 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Страница 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Страница 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Страница 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Страница 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Страница 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Страница 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Страница 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Страница 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Страница 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Страница 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...