Maintenance
6
Temperature Accuracy Test
Required Tools and Materials
You will need a calibrated temperature source (e.g., Fogg System Co. model TP-
400 or equivalent).
Procedure
To perform a temperature accuracy test, follow these steps in the indicated order:
1.
Set the calibrated temperature source to exactly 37º C and plug it into
connector T1.
2.
Start the System and proceed to the Standby screen.
3.
Observe the patient temperature displayed on the screen. It should
indicate 37º C ± 0.2º C.
4.
If the displayed temperature is above or below the indicated range,
contact your ZOLL service representative.
5.
If the displayed temperature was within the indicated range, the test is
complete.
Clean Console and Condensate Pan
Required Tools and Materials
You will need a soft, lint-free cloth and a solution of mild detergent and water to
perform these tasks.
Procedure
To clean the System console, follow these steps in the indicated order:
1.
Assure that the System has been switched off and the power cord has
been unplugged.
2.
Clean the exterior of the console using a soft cloth dampened with a mild
detergent and water mixture. Never use solvents or abrasive cleaners on
the console. Avoid vigorous scrubbing, especially on the front surface of
the display.
3.
Wipe down the outside surfaces with a water-dampened cloth to remove
remaining spots or residue.
4.
Slide out and remove the condensate collection pan from under the front
of the console (see the following figure).
600479-001 Rev 2 136/147
Содержание thermogard xp
Страница 1: ...Operation Manual 600479 001 Rev 2 1 147...
Страница 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Страница 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Страница 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Страница 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Страница 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Страница 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Страница 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Страница 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Страница 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Страница 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Страница 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...