Operation
23
T2 Probe Disconnection/Reconnection
If the User connects a T2 probe at any time during the operation of the System
(in STANDBY or RUN), the System will assume that the use of the T2 probe is
desired. If the T2 probe is then disconnected the System will alarm – see below.
If the User disconnects a T2 Probe in STANDBY, the System will NOT assume
that this is intentional. When the System is placed into RUN WITHOUT a T2
Probe present, the User will again see the Warning as described above.
Accidental Disconnection T1/T2 Probe
Disconnection of the T1 probe during RUN results in
an Alarm. Press the knob to silence the alarm. The
System moves to STANDBY. Treatment cannot
continue until the T1 probe has been replaced..
Note that the patient temperature is displayed as
“---“ and the yellow warning banner covers the bath
temperature display.
Absence or disconnection of the T1 probe during
STANDBY results in a warning display WITHOUT a
persistent audible alarm. The system CANNOT
enter RUN until a T1 probe has been connected.
Note that the patient temperature is displayed as
“---“ and the yellow warning banner DOES NOT
cover the bath temperature display.
Disconnection of the T2 probe during RUN results in
an Alarm. Press the knob to silence the alarm. The
System moves to STANDBY. If you attempt to
return to RUN without reconnecting the T2 probe,
you will be asked to verify your intention.
Note that the patient temperature will be displayed
correctly if the T1 probe remains connected. The
yellow warning banner covers the bath temperature
display.
600479-001 Rev 2 54/147
Содержание thermogard xp
Страница 1: ...Operation Manual 600479 001 Rev 2 1 147...
Страница 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Страница 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Страница 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Страница 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Страница 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Страница 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Страница 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Страница 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Страница 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Страница 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Страница 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...