Operation
7
Table 4-1. Setup Sensor Checks
Air Trap
There is no Start-up Kit installed or there is a large amount of air in the
Start-up Kit chamber.
Roller Pump Lid
The clear plastic roller pump lid is not closed properly.
Check Prime Switch
The Prime Switch is being depressed (as when you use it to prime the
system) or has been jammed in the ‘on” position.
Coolant
The coolant level is low. Top up the coldwell with coolant.
If any of these sensor states are not correct you will see the “System Setup” and
the problem sensor state will be highlighted in red (see Figure 4-1 above). Once
you rectify the identified problems, the System will progress to STANDBY.
NOTE: Change in Design – Patient Temperature Probes
In earlier versions, the System would not enter STANDBY without at least a
Primary Patient Temperature Probe (T1) connected to the System. For XP
Systems, the System will enter STANDBY without a T1 probe being
connected; however, the System will NOT enter RUN without at least the
Primary Patient Temperature Probe (T1) connected to the System (see below:
T1/T2 Behavior -
New
).
STANDBY
Figure 4-2. System Operating Display - Standby
In STANDBY you can interact with the full user interface of the System. You can
select the target temperature, the rate and the treatment mode.
Patient temperature alarms are active in STANDBY if the primary patient
temperature probe (T1) is connected.
The coolant bath temperature graph is active in STANDBY (a new feature
compared to earlier versions of the System).
600479-001 Rev 2 38/147
Содержание thermogard xp
Страница 1: ...Operation Manual 600479 001 Rev 2 1 147...
Страница 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Страница 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Страница 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Страница 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Страница 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Страница 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Страница 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Страница 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Страница 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Страница 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Страница 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...