42 |
Compress Device
Surgical Technique
Spindle Implantation
(cont.)
Realignment of Spindle
If the spindle alignment needs to be changed,
disassociate the provisional and replace the
compression cap with firm clockwise manual rotation
(Figure 81 and 82).
Reintroduce the nut driver and in a counterclockwise
direction and make one full turn (360 degrees) to
loosen the nut. Change the rotation of the spindle
(Figure 83) and re-start the locking process (beginning
with Figure 71).
Pin Drilling and Insertion
The Pin Placement Table specifies which array of holes
may be used based on the centering sleeve that was
previously selected (Figure 84).
Pin Placement Table
Spindle
Size
Inner
Array
Middle
Array
Outer
Array
30 mm
13-14 mm
Sleeve
–
–
38 mm
13 mm
Sleeve
13-18 mm
Sleeve
13-22 mm
Sleeve
44 mm
13 mm
Sleeve
13-20 mm
Sleeve
13-27 mm
Sleeve
Elliptical
(37x49 mm)
13-15 mm
Sleeve
13-18 mm
Sleeve
13-28 mm
Sleeve
Figure 82
Figure 83
Figure 81