15 |
Compress Device
Surgical Technique
Firm Contact
Figure 24
Figure 25
Anchor Plug Implantation
With the drill guide properly aligned with the anchor
plug, remove the three drill bits and position the
anchor plug into the femoral canal. Make certain
that the cross bar is in firm and even contact with
the femoral osteotomy site and that the centering
sleeve provisional is snug within the canal (Figure 25).
Placement of the anchor plug within the medullary
canal is non-directional; location is typically dictated
by the best exposure and ease of access to the drill
guide.
Anchor Plug
(cont.)
When the knobs are tight, all three drill bits should
slide freely through the alignment holes (Figure 24).
Note:
If the drill bits do not slide freely, loosen the
top two adjustment knobs and begin the tightening
process again.