40 |
Compress Device
Surgical Technique
Spindle Implantation
(cont.)
Standard Distal Femoral Trialing
Place the 5 cm taper adapter provisional over the
spindle taper and assemble the additional provisionals
to achieve the correct construct length (Figures 76
and 77).
Reduced Resection Distal Femoral Trialing
When trialing for the Compress 7 cm, 8.5 cm or 10 cm
reduced resection implants (proximal or distal) (Figure
78), it is necessary to place the Compress 5 cm taper
adapter provisional with the threaded connection (32-
481075) over the exposed spindle taper (Figure 79).
Place the Compress reduced resection provisional
over the no face adapter provisional until the magnets
inside the blue provisional engage with the taper
adapter provisional (Figure 80).
Figure 78
Figure 77
Figure 76
Содержание OSS Compress
Страница 1: ...OSS Orthopedic Salvage System Compress Device Surgical Technique ...
Страница 2: ......
Страница 43: ...41 Compress Device Surgical Technique Figure 79 Figure 80 ...
Страница 58: ...56 Compress Device Surgical Technique Notes ...
Страница 59: ......