9 |
Compress Device
Surgical Technique
A
B
13 mm only
Anchor Plug
Instrumentation Assembly
Important:
There are two distinct anchor plug
holders in the Compress instrumentation. The first
(Figure 11A) is used without a centering sleeve
provisional when the final reamer at the resection
level is 12 mm. It is also used with centering sleeve
provisionals when the final reamer measures 14
mm through 28 mm.
The second anchor plug holder (Figure 11B) is used only
when the final reamer at the resection level measures
13 mm. This 13 mm anchor plug holder is used without
a centering sleeve provisional.
Slide the anchor plug holder through the cross bar
(Figure 12).
Figure 11
Figure 12