32 |
Compress Device
Surgical Technique
Spindle
(cont.)
Important:
There are two distinct reamer pilots
in the Compress instrumentation. The first (Figure
59A) is used without a centering sleeve provisional
when the final triple reamer diameter at the
resection level measures 12 mm. It is also used with
centering sleeve provisionals when the final triple
reamer diameter measures 14 mm through 24 mm.
Important:
The 13 mm reamer pilot (Figure 59B) is
used only when the final triple reamer diameter at
the resection level measures 13 mm. The reamer
pilot is used without a centering sleeve provisional.
Place the face reamer blade over the two retaining
tabs located on the reamer body and secure with
a clockwise turning motion. Insert the reamer pilot
into the face reamer and tighten the side set screw
(Figure 60).
A
B
Figure 59
Figure 60