4 |
Compress Device
Surgical Technique
Resection
Utilize the Compress Resection Chart on page 2 to
determine the correct implant components needed to
recreate the residual femur.
To achieve a proximal or distal femoral resection,
begin by making a resection mark and a rotational
alignment mark with a cautery device or methylene
blue (Figure 3). Using a freehand transverse cut, resect
the femur at the reference resection mark.
Figure 3
Figure 4
Figure 5
No Face Adapter Provisional
Construct
Following the transverse resection, it is recommended
that the no face adapter provisional be attached to
segmental provisionals in order to ascertain some
semblance of the construct length (Figures 4 and 5).
This is suggested at the onset of the surgery because
once the holes for the transverse pins are drilled
and the anchor plug is secure within the canal, no
additional bone may be resected without dramatically
altering the original replacement length.
Note:
If preparing the bone for use with the
Compress reduced resection implants, utilize the
no face adapter provisional with the threaded
connection rather than the quick connect
(32-481114).
For use with the Reduced Resection Implants
Part Number 32-481114
For use with the Standard or Short Implants