Service Instructions VISUCAM
lite
Issue: 22.09.2003
SM-30-4044-A1-en
Replaces issue:
Page
22
of 41
Opening and closing the control base
•
Removing the camera head including
instrument base.
- Disconnect the cable for the camera
head.
- Loosen the screws in the toothed
rack.
(2.5mm hex socket head)
- Push out the camera head including
instrument base to one side.
- Place the camera head including
instrument base on the table.
•
Opening the control base
- Put the base including headrest to one
side.
- Remove 8 x 3 mm hex socket head
screws.
- Remove 4 x Phillips screws from
the connector panel.
- Remove the base cover and lay it
aside.
Caution:
Take care not to tear off the fixation lamp cable. (see also replacing the headrest!).
•
Close the base in reverse order.