Replacing the fan cooling the lamp housing
Note:
Fan blows outward. Observe markings on the fan!
•
Open the camera head (see:
Opening and closing
the camera head
).
•
Disconnect the fan cable from X45 on the lamp
control PCB.
•
Screw off the defective fan.
•
Mount the new fan and connect it to X45
on the lamp control PCB.
•
Close the camera head
(see:
Opening and closing the camera head
).
•
Switch on VISUCAM
lite
and check its functions.
Replacing the temperature sensor
•
Disconnect VISUCAM
lite
from line power.
•
Open the camera head (open left half only)
(see:
Opening and closing the camera head
).
•
Disconnect the cable of the temperature sensor
from X46 on the lamp control PCB.
•
Screw off the defective temperature sensor.
•
Mount the new temperature sensor and connect
it to X46 on the lamp control PCB.
•
Close the camera head
(see:
Opening and closing the camera head
).
•
Switch on VISUCAM
lite
and check its functions.
Replacing Shutter
•
VISUCAM
lite
vom Netz trennen.
•
Disconnect VISUCAM
lite
from line power.
•
Open the camera head (open left half only)
(see: Opening and closing the camera head).
•
Disconnect the cable of shutter from X47 on the lamp
control PCB.
•
Screw off the defective shutter. Shutter wing is still on
place.
•
Mount the new shutter and connect it to X47 on the lamp
control PCB.
•
Close the camera head
(see: Opening and closing the camera head).
•
Switch on VISUCAM
lite
and check its functions.
Shutter
X47
Содержание VISUCAM lite