Service Instructions VISUCAM
lite
Issue: 22.09.2003
SM-30-4044-A1-en
Replaces issue:
Page
24
of 41
Components of the control base
Notes on the removal and the installation of the base components:
All interchangeable components can be removed and installed without difficulty.
The continuous external saving of images and patient data has to be done by the
customer itself. Both a hard disk with an operating system and a VISUCAM
lite
program
and a blank hard disk are available.
Exchanging power supply
For exchanging the power supply there is an Allen key 3/32 inch included in the Service kit.
red
+5V
black
ground
yellow
+12V
CD-ROM / drive, writer
Inductive approximation
switch
Med PIII Motherboard
12VDC fan
Connection
field
Hard disk
GIF instrument interface
+5V, +12V
power supply
Line module
PS2 Interface
Mouse, Keyboardr
Power input
Содержание VISUCAM lite