Service Instructions VISUCAM
lite
Issue: 22.09.2003
SM-30-4044-A1-en
Replaces issue:
Page
21
of 41
Replacing the VISUCAM
lite
control PCB
Note:
When exchanging the control PCB, re-enter and save the serial number of the
instrument. Also enter and save the instrument type (see
Data input fields
).
•
Open the camera head (see:
Opening
and closing the camera head
).
•
Unplug all cable connections.
•
Screw out 4 fastening screws and remove
the PCB.
•
Insert the new VISUCAM
lite
control PCB
and secure it in position.
•
Re-connect the cables.
•
Close the camera head (see:
Opening
and closing the camera head
).
•
Switch on VISUCAM
lite
and check its
functions.