Service Instructions VISUCAM
lite
Issue: 22.09.2003
SM-30-4044-A1-en
Replaces issue:
Page
11
of 41
Separating the camera head from the instrument
base
•
Open the camera head
(see:
Opening and closing the camera head
).
•
Disconnect the cable for the camera head
X14
,
X20
from the connector panel.
•
Disconnect the 2 cables leading to the
instrument base from the VISUCAM
lite
control
PCB. Remove the cable ties.
•
Push the cables through the connecting holes.
•
Remove the securing ring.
•
Close the camera head (see:
Opening and closing the camera head
).
•
Pull the camera head from the pin and lay it aside.
•
Mount the new camera head accordingly.
•
Switch on VISUCAM
lite
and check its functions.
Replacing the VISUCAM
lite
control cable
•
Open the camera head (see:
Opening and
closing the camera head
).
•
Disconnect the control cable from the
connector panel.
•
Disconnect the control cable from the
VISUCAM
lite
control PCB, the cameras
control PCB and the lamp control PCB.
•
Remove cable ties and cable clamps from
the cable.
•
Route the new VISUCAM
lite
control cable
as per wiring diagram and connect it.
•
Close the camera head (see:
Opening and
closing the camera head
).
•
Switch on VISUCAM
lite
and check its
functions.
Securing ring
Control cable
Control PCB
VISUCAM
lite
Cameras control PCB
Содержание VISUCAM lite