26
G-30-1702-en
OPMI® Pentero® Software Release 2.2
Issue 2.0
Printed on 17. 03. 2008
What to do in an emergency
1
Failure of
illumination
- changing the xenon lamp:
•
Open flap (1)
•
Change the xenon lamp by pulling grip (8)
2
Failure of the
zoom function
:
•
Manually adjust the magnification using zoom knob
(2). If the motorized zoom function becomes active of
its own accord (e.g. travels to the stop), set emergency
switch (7) to the emergency mode (position 2).
3
Failure of the
focusing function
:
•
Manually adjust the working distance using focusing
knob (3). If the motorized focusing function becomes
active of its own accord (e.g. travels to the stop), set
the emergency switch to the emergency mode (posi-
tion 2)
4
Some of the
magnetic brakes
are blocked:
•
Switch off power switch (4). As soon as the blue
screen appears (approx. 10 sec), switch the system
back on.
The OPMI functions (zoom, focus, light and magnetic
brakes) are available again after approx. 15 seconds.
The computer and touchscreen, however, are disa-
bled.
If the magnetic brakes are still blocked:
•
Hold the microscope on its body (not on the handgrips)
and position it manually by overcoming the braking ef-
fect.
5
Failure of the
touchscreen
:
•
Do not under any circumstances touch the touch-
screen, since this can result in changes to settings and
parameters. Zoom, focus, illumination and brakes can
still be operated.
6
Error messages in the
data injection system
:
•
System errors are displayed in the microscope's inte-
grated data injection system. You can delete these
messages by acknowledgement using the joystick of
the right handgrip (pushbutton) or the touchscreen.
Содержание OPMI Pentero
Страница 1: ...OPMI Pentero Software Release 2 2 Brief instructions G 30 1702 en Issue2 0 Printed on 17 03 2008...
Страница 7: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 7 3 4 5 6 5 6...
Страница 25: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 25 BNC 1 Y C 2...
Страница 29: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 29 1 2 3...
Страница 39: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 39...