G-30-1702-en
OPMI® Pentero® Software Release 2.2
Issue 2.0
Printed on 17. 03. 2008
9
Balancing the system
Warning
!
Make sure to observe the instructions in the chapter "Balancing the
system" of user manual G-30-1458.
For safety reasons, the system must only be used when correctly bal-
anced. Despite the autobalance function, it may happen in exceptional
cases that the surgical microscope is not correctly balanced.
With an incorrectly balanced system, brake release may lead to uncon-
trolled movements of the suspension system. For this reason, the bal-
ancing procedure and the subsequent test must not be performed above
the patient and only at a safe distance from other persons and devices.
•
Hold the microscope tightly at both handgrips.
•
Unlock the magnetic brakes (AB) and swing out the microscope until
indicator (1) is centered in the blue area (2). Then you can perform the
"Autobalance Complete System" procedure.
SB
AB
Fine adjustment for mouth switch
G-30-1458
PHOTO
DRAPE
FREEZE
Autobalance Complete System
MENU
START
START
START
Autobalance Microscope
Drape Compensation
Autobalance
30. 01. 04
17:56
User:
XXXXXXXX
Patient:
XXXXXXX
Recorder Capacity:
Int:.......% Ext:.........%
USER
VOICE CTRL
DICTATION
TELEPHONE
FULL SCREEN
CLOSE
1
2
CONFIG
PAT-FILES
Fine adjustment mouth switch
-3
-
+
REC START
STOP
AUTOBALANCE
Содержание OPMI Pentero
Страница 1: ...OPMI Pentero Software Release 2 2 Brief instructions G 30 1702 en Issue2 0 Printed on 17 03 2008...
Страница 7: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 7 3 4 5 6 5 6...
Страница 25: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 25 BNC 1 Y C 2...
Страница 29: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 29 1 2 3...
Страница 39: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 39...