18
G-30-1702-en
OPMI® Pentero® Software Release 2.2
Issue 2.0
Printed on 17. 03. 2008
Attaching documentation / coobservation equipment
•
Loosen knurled ring (1).
•
Remove dust cover (2) and keep it in a safe place.
•
Insert accessory module (3) in the mount of the image exit port and
carefully turn it until the guide projections fit into the grooves. Push the
accessory module into the mount as far as it will go.
•
Screw knurled ring (1) onto accessory module (3) and firmly tighten it.
The type of coobservation (lateral image exit ports: left/right, or opposite
image exit ports: face to face) can be configured at the touchscreen
(CONFIG/OPMI/TUBE).
The sliding mirror has two positions:
Left/Right:
The light is directed to the lateral image exit ports.
Face to Face
: The light is directed to the tube mount at the back.
The position of the sliding mirror can also be switched manually using
knob (4).
Warning!
Check that all modules are securely seated. Firmly tighten se-
curing screws (8) and (9)! Balance the surgical microscope, page 9.
G-30-1458
3
1
2
4
2
3