Setting the eyepieces (compensation for ametropia)
Bring the surgical microscope into its starting position within the focusing range. Adjust the minimum magnification on the surgical
microscope. Bring the surgical microscope into the position required. Adjust your interpupillary distance on the binocular tube. Adjust
your prescription on the eyepieces. Please note that instrument myopia can occur.
Emmetropes
adjust the diopter scale to 0 diopter
(D).
Eyeglass wearers
(who perform surgery wearing their
eyeglasses) adjust the diopter scale
to 0 diopter (D).
Ametropes
(who know their prescription and
perform surgery without wearing
their glasses) adjust the diopter sca-
le to their prescription.
Ametropes
(who do not know their prescription and perform surgery without wea-
ring their glasses)
Adjust both eyepieces on the main microscope to +8 D.
Adjust both eyepieces on the assistant's microscope to +8 D (if applicable).
Eyepieces without reticle:
Eyepieces with reticle:
Remove the binocular tube and the
eyepieces from the microscope
body and point them at a distant ob-
ject
*)
, i.e. use them like a pair of bi-
noculars. The image of the object is
still unsharp.
Turn the diopter scale on the eyepiece
slowly clockwise until the reticle is
imaged sharply.
Turn the diopter scale of one eye-
piece slowly clockwise until the ob-
ject is imaged sharply. If necessary,
repeat this procedure three times
and take the average of the rea-
dings. Adjust the second eyepiece
in the same manner.
Focus the surgical microscope on the
object. The reticle and the object must
be in focus at the same time.
Mount the tube and eyepieces on
the microscope body and tighten
the securing screw firmly.
Adjust the second eyepiece in the same
manner until the object is seen sharply.
Adjust the eyecups in such a way that you can see the full field of view. Adjust maximum magnification on the surgical microscope
and focus on the object. Adjust the working magnification required. When the magnification is changed, the focal plane is retained,
but the depth of field changes.
Note: Enter the prescription values in the relevant user profile.
*) Warning!
Never use the sun as the distant object!
Содержание OPMI Pentero
Страница 1: ...OPMI Pentero Software Release 2 2 Brief instructions G 30 1702 en Issue2 0 Printed on 17 03 2008...
Страница 7: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 7 3 4 5 6 5 6...
Страница 25: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 25 BNC 1 Y C 2...
Страница 29: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 29 1 2 3...
Страница 39: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 39...