Malfunctions during measuring operations
8-4
61211-8140502
METROTOM 1500 G3 1.4
Malfunctions during measuring operations
The following table helps you identify and eliminate malfunctions that
occur during measuring operations. A malfunction can have different
causes. One or more measures may be necessary to eliminate a certain
cause. If two measures are specified, this does not mean that you have
to perform both of them to fix the problem.
Malfunction
Cause
Action
Check
No travel movement
possible
The drives are off
–
Press the two
buttons
of
the loading door simulta-
neously.
–
Then press the
button
for
the drives.
The indicator lamp
of the switch
lights up.
The
EMERGENCY STOP
button
is locked
Unlock the
EMERGENCY
STOP
button:
–
Turn the button slightly.
The button pops
out a few millime-
ters.
Collision, for example
workpiece with X-ray
tube
–
Eliminate the cause of the
collision.
–
Reactivate the drives.
–
Reinitialize the X-ray tube
if necessary.
Wrong operating mode
set.
The
key switch
is in one
of the following posi-
tions:
,
.
–
Turn the
key switch
to the
following position.
A CT measurement is
running.
–
Wait until the CT mea-
surement has been com-
pleted.
The voltage in the X-
ray tube cannot be
increased.
Too long pause between
two measurements.
–
Perform a warm-up.
No initialization has
been carried out.
–
Perform a complete ini-
tialization.
If measuring operation is still not possible, call a ZEISS service engineer
or our support team.
➤
Содержание METROTOM 1500 G3
Страница 8: ...6Table of contents 6 61211 8140502 METROTOM 1500 G3 1 4...
Страница 18: ...Warranty 1 6 61211 8140502 METROTOM 1500 G3 1 4...
Страница 72: ...Start up checklist 6 14 61211 8140502 METROTOM 1500 G3 1 4...
Страница 104: ...Inspection measures 9 10 61211 8140502 METROTOM 1500 G3 1 4...
Страница 110: ...2Glossary 2 61211 8140502 METROTOM 1500 G3 1 4...
Страница 120: ......