35 – LSM 880 Training Guide
Aligning AiryScan
5.
Allow scan to repeat until automatic alignment quality and status reads
‘Good’ OR until AiryScan detector view shows evenly illuminated array (as
shown in example image).
Note: Initially you may see the quality and status readout ‘Out of Range’ or
‘Bad’ while the alignment is running. These are normal readouts during
optimization and should go away once the optimization is complete.
6.
Stop scan.
7.
Click ‘Snap’ to collect image.