32 – LSM 880 Training Guide
Using AiryScan
When we have finished balancing the available signal to fill 50% of the dynamic range, we can further prepare the system for
scanning at the best possible resolution.
1.
Go to Acquisition Mode and under Frame Size, click ‘Optimal’.
Note: Using the Optimal feature is essential for sampling the field of view
correctly for AiryScan. The raw data needs to be sampled at Nyquist (2x) in
order to achieve the best quality image.
Scanning at resolution settings below Nyquist (2x) will result in lower
resolution files – and possibly create artifacts in the processed data.
2.
Set scan speed to Max.