Installation
Daily commissioning
WARNING - GENERAL HAZARD
Before using the device, the user must ensure that it is functioning
correctly and is in proper condition, and the user must observe the
Instructions for use - Operating.
The following inspection measures must be completed before use
each workday:
•
Visual inspection of the housing, exterior labeling, Instructions for
use - Operating, accessories and power cable for damage. All of
these elements must be present. If parts are missing or there is
visible damage, the device may not be used and must be shut
down.
•
The ventilation slots on the housing may not be covered or
blocked!
•
Functional testing of measurement function using the test eye de-
vice.
WARNING - DANGER OF ELECTRICAL SHOCK
Please ensure that the following requirements are fulfilled before each
commissioning, and that they continue to be fulfilled during ongoing
operations:
•
The device is connected with the power cable intended for the device.
If an instrument table is used, the device receives electrical power
through the table.
If a table which has not been authorized by Carl Zeiss is used, the user
bears sole responsibility for the electrical safety of the device.
•
The power plug must be inserted into an outlet with a proper
protective conductor connection.
•
All cables and plugs may only be used if they are free from defects.