Electromagnetic
compatibility
57
EN 20_070_0009I v1.5
Table 3: Guidance and Manufacturers Declaration - Electromagnetic Immunity
The i.Profiler
plus
is intended for use in the electromagnetic environment specified below. The customer or the
user of the i.Profiler
plus
should ensure that it is used in such an environment.
Immunity
testing
IEC 60601
testing level
Compliance
level
Electromagnetic environment -
guidelines
Conducted RF IEC
61000-4-6
Radiated RF IEC
61000-4-3
3.0; 6.0 Vrms
150 kHz to 80
MHz outside ISM
bands
10 V/m
80 MHz to 2,5
GHz
3.0; 6.0 Vrms
150 kHz to 80
MHz
10 V/m
80 MHz to 2,5
GHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the i.Profiler
plus
, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended protection distance
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol.
Immunity from
Proximity field
from wireless
communications
equipment
List of frequencies,
from 9 V/m up to
28 V/m,
PM (18 Hz or 217
Hz), FM 1 kHz
List of frequencies,
from 9 V/m up to
28 V/m,
PM (18 Hz or 217
Hz), FM 1 kHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines
may not apply
in all conditions. Electromagnetic propagation is affected by
absorption and reflections of buildings, objects and people.