Electromagnetic
compatibility
55
EN 20_070_0009I v1.5
Environment of intended uses: Home Healthcare Environment
Table 1: Guidance and manufacturer's declaration - electromagnetic emissions
The i.Profiler
plus
is intended for use in the electromagnetic environment specified below. The customer
or the user of the i.Profiler
plus
should assure that it is used in such an environment.
Emission testing
Accordance
Electromagnetic environment
- guidelines
RF emissions CISPR 11
Group 1
The i.Profiler
plus
uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions according to CISPR 11 Class B
The device or system is suitable for use in all
establishments, including those in the home
and those directly connected to the public
low-voltage network, which also supplies
buildings used for residential purposes. The
device has a power consumption of <200 W.
Emissions from
Harmonics
IEC 61000-3-2
Not applicable
(for 230 VAC)
Class A (for 220 VAC)
Voltage fluctuations/
Flicker emissions IEC
61000-3-3
Complies
The i.Profiler
plus
is suitable for use in all
establishments, including domestic
establishments and those directly connec-ted
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.