56
Electromagnetic compatibility
EN 20_070_0009I v1.5
Table 2: System guidance and manufacturer's declaration - electromagnetic immunity
The i.Profiler
plus
is intended for use in the electromagnetic environment specified below. The customer or user of
the device should ensure it is used in such an environment.
Immunity testing
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidelines
Electrostatic discharge
(ESD) IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV
±15 kV
Floors should be wood, concrete or
ceramic tile. If the floor is covered
with synthetic material, the relative
humidity must be at least 30%.
Electrical fast
transient/burst IEC
61000-4-4
± 2 kV for power supply
lines;
±1 kV for input/
output lines
± 2 kV and
± 1 kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge voltage
IEC 61000-4-5
±1 kV
line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line to line
± 2 kV line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
0 % UT (100% dip in UT)
for 0.5 cycle
for 10 msec and 20
msec
70 % for 500 msec
0 % for 5 s
0 % UT (100% dip in
UT) for 0.5 cycle
for 10 msec and 20
msec
70 % for 500 msec
0 % for 5 s
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
i.Profiler
plus
requires continued
operation during power mains
interruptions, it is recommended
that the
i.Profiler
plus
be powered from an
uninterruptible power supply or a
battery.
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE:
UT is the AC mains voltage prior to applying the test level.