DICOM Gateway
Humphrey Matrix User Manual
2660021138935 Rev. F 2015-11
A-4
Local Application Entity Settings
Institution Name: Typically represents the name of the hospital or clinic.
Enter a total of up to 64 characters and spaces.Station Name: Typically a unique name for
the exam lane or instrument. Enter a total of up to 16 characters and spaces.
Local AE Title: Enter a unique AE Title for the Matrix instrument. This AE Title needs to be
registered with your DICOM system. Input a total of up to 16 characters (text is
case-sensitive). MWL - Today’s Patients automatically searches for patients scheduled for
an exam for this Local AE Title on today’s date (see
MWL - Today’s Patients (Worklist
Patients Scheduled for Today)
on page
5-2
).
Note: Each Matrix instrument must have its own unique AE Title which needs to be
registered with your DICOM system.
DICOM Port: Enter the Local Port Number for DICOM Gateway that your DICOM system
(e.g., FORUM) connects to.
Remote Application Entities
MWL and Storage Provider AE Title: Enter your DICOM System’s MWL and Storage Provider
Application Entity Title. Input a total of up to 16 characters (text is case-sensitive).
MWL and Storage Provider Host Name: Enter your DICOM System’s MWL and Storage
Provider Host Name or IP Address.
MWL and Storage Provider Port: Enter your DICOM System’s MWL and Storage Provider
Port number.
Note: The other Remote Application Providers (Storage Commitment Provider, etc.) are
disabled in this software release.
Once you have all the fields completed, click the Test button. A green status indicator light
(
) will appear to the left of MWL and Storage Provider and the Connection Test Details
will report “Passed” for a successfully configuration (see
Figure A-3
below). A yellow
status indicator light ( ) may be visible during the connection test if it takes time to
complete the connection test. If a red status indicator light ( ) appears, it means the
configuration failed the connection test—see the text in Connection Test Details for more
information.
Note: When you change a field, the status indicators turn gray until you run another test.
Содержание Humphrey Matrix 800
Страница 1: ...User Manual Humphrey Matrix Model 800...
Страница 26: ...Introduction Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 1 20...
Страница 84: ...Viewing and Printing Tests Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 6 18...
Страница 108: ...Networking Configuration Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 8 18...
Страница 150: ...OfficeMate PMS Instructions Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 B 2...
Страница 156: ...Error Codes Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 D 4...
Страница 157: ......