Error Codes
Humphrey Matrix User Manual
2660021138935 Rev. F 2015-11
D-2
197
Are you sure you want to reassign
these NUMBER_OF_TESTS tests?
If the user highlighted the tests and
selects the Reassign Test option.
The user decides to reassign the
selected tests to another patient.
202
The character
AN_INVALID_CHARACTER cannot
be used as a text delimiter.
While in the Save As Dialog box, the
user attempts to assign a delimiter
which is invalid. Invalid characters
include: (the double quote, ")
Specify a different delimiter. Reboot
and try again. If persistent, use a
predefined delimiter and report error
to developers.
238
This operation is not available.
Because the network host was not
available the file write test could not
be attempted.
Verify that the Matrix and the shared
PC are connected to the network. If
both are connected make sure file
and print sharing are enabled on the
PC are turned on and that the
appropriate folders and printers have
been shared.
239
Please enter a name for the printer.
Give the printer selected a reference
name.
240
Please enter a valid hostname or IP
address
Ask the operator to enter a shared
network computer host name or ip
address (computer1.cznet.org,
128.115.68.2). This message is no
longer used.
Enter in a hostname
(mypc.domain.com) or IP address
(128.68.24.2).
241
Please select a path for the printer.
Asks the operator to specify the
location of the network shared
printer. This message is no longer
used. The printer is now selected
graphically.
Enter the path to the network shared
printer
(//mycomputer.domain.com/HP6920)
.
262
The device is busy and cannot safely
be removed.
The storage device is currently being
used for read or write operation.
Wait for the operation to complete
before trying to remove the storage
device. If this situation persists, try
rebooting the Matrix. The period of
time required to wait will depend on
the operation performed. Backup or
export of a large number of patients
and tests will take a long time.
513
Overview Software Error! Function
not available.
Software implementation error.
An error has occurred during SW
processing. Report event to Service.
New software may be required.
515
Overview plot Software Error!
Function not available.
Software implementation error.
An error has occurred during SW
processing. Report event to Service.
New software may be required.
540
Test parameter SW Error! Test
cancelled. Function not available.
SW implementation error.
This error indicates a defect in the
SW processing has occurred. Contact
Service.
552
Database backup not found on
media.
The database backup could not be
found.
Make sure the media is connected to
the device and database files are on
the selected storage media.
For a restore, the files are placed in a
directory named SN<serial number
of the Matrix>.
Содержание Humphrey Matrix 800
Страница 1: ...User Manual Humphrey Matrix Model 800...
Страница 26: ...Introduction Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 1 20...
Страница 84: ...Viewing and Printing Tests Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 6 18...
Страница 108: ...Networking Configuration Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 8 18...
Страница 150: ...OfficeMate PMS Instructions Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 B 2...
Страница 156: ...Error Codes Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 D 4...
Страница 157: ......