Humphrey Matrix User Manual
2660021138935 Rev. F 2015-11
Error Codes
D-3
576
Printer Error! Unable to print System
Information!
Numerous printer errors:
• Not connected
• No paper in tray
• Non-approved printer
• No ink
Verify the printer attached to the
system is an approved printer.
Perform the print sample test from
the Diagnostics screen.
If problem persists, cycle the printer
power or Shut Down and reboot the
Matrix instrument.
If problem still persists, try a different
approved printer.
584
Unable to retrieve login credentials
for __________________
Try again. Check network.
593
Failure to export exam to DICOM
archive
Check DICOM archive. Check DICOM
configuration in System Settings.
619.2
Critical Error! Factory calibration
needed. Select 'Shut Down' and then
cycle power to restart your
instrument and restore normal
operation. If problem persists,
contact Customer Service. See Help
F6.
User attempted to perform the
pattern test 'Set Pattern' from the
Advanced Diagnostics screen. This
operation could not complete
because their system needs to be
factory calibrated.
SW detected a problem that requires
Shut Down and restart to clear.
Service will be needed if this issue is
chronic and prevents normal
operation of the system.
619.3
Critical Error! Factory calibration
needed. Select 'Shut Down' and then
cycle power to restart your
instrument and restore normal
operation. If problem persists,
contact Customer Service. See Help
F6.
User attempted to perform an
'Ambient Light Level' check from the
Advanced Diagnostics screen. This
operation could not complete
because their system needs to be
factory calibrated.
SW detected a problem that requires
Shut Down and restart to clear.
Service will be needed if this issue is
chronic and prevents normal
operation of the system.
619.4
Critical Error! Factory calibration
needed. Select 'Shut Down' and then
cycle power to restart your
instrument and restore normal
operation. If problem persists,
contact Customer Service. See Help
F6.
User attempted to perform a
'Monitor Brightness' check from the
Advanced Diagnostics screen. This
operation could not complete
because their system needs to be
factory calibrated.
SW detected a problem that requires
Shut Down and restart to clear.
Service will be needed if this issue is
chronic and prevents normal
operation of the system.
702
Unable to connect to DICOM server.
Check DICOM archive. Check DICOM
configuration in System Settings.
704
One or more patients could not be
imported.
Some of the patients could not be
imported from the serial EMR
application.
Check the EMR connection and
setup.
Содержание Humphrey Matrix 800
Страница 1: ...User Manual Humphrey Matrix Model 800...
Страница 26: ...Introduction Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 1 20...
Страница 84: ...Viewing and Printing Tests Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 6 18...
Страница 108: ...Networking Configuration Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 8 18...
Страница 150: ...OfficeMate PMS Instructions Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 B 2...
Страница 156: ...Error Codes Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 D 4...
Страница 157: ......