Quick Reference Guide
HFA3 Quick Reference Guide
2660021166131 Rev. A 2018-11
5-2
Test Setup
1. For an existing patient the previous test that was used will automatically display.
Otherwise, the 24-2 SITA FASTER is the default test.
2. If needed, select a different test from the Test Profile menu. The new test profile,
pattern, and strategy now displays.
3. If a specific test is not in the Test Profile menu, swipe down on the page and use the
Test Parameters menu to select new test parameters.
4. Use the Perform test on drop-down menu to select which eye to test: OD (right), OS
(left) or Both. Default is both.
5. To add notes select Add Note.
6. Change the default Reports by selecting or deselecting one or more Reports from the
drop-down menu.
7. Enter IOP (intraocular pressure) and Visual Acuity if desired.
8. Confirm test information and select the Next button to proceed.
Trial Lens Setup
The Auto Trial Lens Correction (AutoTLC) automatically adjusts the spherical correction to the Trial
Lens value. This feature is only available on Model 860.
For all other models insert manual trial lens into the holder:
1. Move the trial lens holder into an upright position from its storage position in the
bottom of the bowl (A).
2. Place the spherical lens in the slot closest to the patient and the handle to the right for
the OD eye and to the left for the OS eye (B).
A
B
C