Specifications
HFA3 Instructions for Use
2660021166131 Rev. A 2018-11
12-2
Storage (Unpacked)
• Temperature: –10 to +55°C
• Humidity: 10% to 95%
• Atmospheric Pressure: 700 to 1060 hPa
Operation
• Temperature: +10 to +40°C
• Relative Humidity: 30% to 90%, excluding condensation
• Atmospheric Pressure: 700 to 1060 hPa
• Altitude: Up to 3000 m above sea level
• Max. wet bulb: 78°F (26°C)
WARNING: If any part of the system is reconfigured or replaced or if any external
devices are connected to the instrument, the operator must ensure that the
complete system continues to comply with the requirements defined in IEC
60601-1.