HFA3 Instructions for Use
2660021166131 Rev. A 2018-11
4-21
Stimulus Value, Speed, and Visual Field settings cannot be altered during the test. SSA test
parameters are III 4 e (size and intensity), 4 degrees (4°)/sec (speed), White (color).
Note: Select Cancel at any time to cancel test results.
To run a preset Kinetic Test:
1. In the Patient screen, go to Test Parameters and choose an available preset test pattern from
the Test Pattern
drop-down menu.
2. Press Next. The Kinetic Test screen displays the Live Eye monitor and the test pattern with
preset test points.
3. Choose between two test modes:
• Select Automatic to run all of the test points automatically.
• Select Step by Step to run each stimulus one at a time.
Note: If desired, switch between Automatic and Step by Step at any time during the test or when
the test is paused.
Automatic Mode
A. Select Start Test.
B. Each stimulus will automatically move along the line to the center. Coordinates are
displayed above the test pattern.
C. Select Pause to stop the test and choose from the following options:
• Select Restart Test to delete current test responses and restart the test. Confirm or
cancel this selection.
• Select Return to Test to continue with the test.
Step by Step Mode
A. Select Start Test to test the first point. The Pause option is available while the test is
running.
B. Select Test Next Point to test each additional point. The Restart Test
option is also
available.
4. During testing, Points Completed is continuously updated and isopter lines are drawn between
completed points. Total Points are also shown.
5. Retest a point by selecting an already tested stimulus, then select Retest Point. This option is
available only after all points are tested.
Note: A message will display at the end of testing if points were missed during the test (no response
made). The option to retest the points will be provided.
6. Select Review Results when testing is complete to test the other eye.
Note
See the feature
.
Review Results
Same as Manual Kinetic
.
Содержание Humphrey Field Analyzer 3
Страница 6: ...HFA3 Instructions for Use 2660021166131 Rev A 2018 11 vi...
Страница 94: ...HFA3 Instructions for Use 2660021166131 Rev A 2018 11 Data Tests Reports Go to Contents 6 18...
Страница 110: ...Maintenance Go to Contents HFA3 Instructions for Use 2660021166131 Rev A 2018 11 8 6...
Страница 155: ......