Corona process | User Manual
Maintenance
Corona process
•
Shut down the control software and disconnect the power supply to the
Corona process
measuring system.
•
Perform the activities described below at an ESD workplace.
•
Have the following spare parts ready:
– Spare lamps, order no. 000000-2219-967
Note
Always replace both halogen lamps together if possible.
•
Have the following tools and accessories ready:
– The tools belonging to the calibration device for the
Corona process
ȩ
Phillips screwdriver PH2 (for older device versions)
ȩ
Open-end wrench, 8 mm A/F
ȩ
Screwdrivers, TX10 and TX25
•
Additionally required tools and accessories:
– Open-end wrench, 22 mm A/F
– Allen key, 6 mm A/F
– Cotton gloves
1
Removing the system cable
Loosen the union nuts of the connected plugs by hand (or with a 22 mm
open-end wrench) and pull the cables out of the connecting sockets on the
measuring system.
Fig. 17
3
2
1
Removing the system cables
"ETHERNET" connection
"Digital IN OUT" connection
"POWER 12 – 24 V DC" connection (power supply)
Requirements
Procedure
1
2
3