
46
Corona process | User Manual
Maintenance
Corona process
Start the
OSISManagementConsole.exe
file under:
⇒
C:\Program Files (x86)\ZEISS\InProcess.
If the
Corona process
measuring system has not yet been configured, select
it for configuration under the menu option
New
.
Fig. 27 Selecting a measuring system
3
Selecting the lamp to be calibrated
Right-click on
dev Corona process
and select
calibration of the default
lamp
or
calibration of the spare lamp
.
Fig. 28 Selecting a lamp
The initialization of the
Corona process
may take a few seconds.