Removing the base of the device from the device cover
Unscrew and remove the four hexagon screws
1
from the housing with
an 8 mm open-end wrench. (Phillips screws may be installed in older device
versions. Loosen Phillips screws with a PH2 Phillips screwdriver.)
Carefully lift the base of the device
2
by the carrying handle
3
out of the
device cover
4
. Make sure that you do not feel any resistance while doing
so.
Turn the base of the device upside down (180°). The interior assemblies are
now located face up and the carrying handle is located face down. Carefully
deposit the base on an ESD workplace.
Fig. 19
3
1
2
4