OPERATION
Axio Scope.A1
Illumination and Contrasting Method
Carl Zeiss
M60-2-0007 e 05/08
71
(4) Adjusting the vertical stop on the
condenser carrier
•
Loosen the setscrew (
625H625H
Fig. 4-2/
1
) on the vertical
stop with the ball-head screwdriver SW 3.
•
Focus the sample with the focusing drive.
•
Close the field diaphragm and adjust the
condenser vertically (
626H626H
Fig. 4-2/
2
) until you get a
sharp image.
•
Perform the vertical adjustment on the
condenser by small increments only to avoid
lifting out the sample.
•
Tighten the setscrew (
627H627H
Fig. 4-2/
1
) on the vertical
stop.
(5) Adjusting the vertical stop on the
focusing drive
•
Turn the clamping lever (
628H628H
Fig. 4-3/
1
) of the stop
towards the pin stop.
•
Move the stage with the aid of the focusing
drive (
629H629H
Fig. 4-3/
2
) to the highest position
allowable without risking colliding with object
carrier or objectives.
•
Now press the clamping lever down until the
stop engages again.
Fig. 4-2
Adjusting the vertical stop on the
condenser carrier
Fig. 4-3
Adjusting the vertical stop on the
focusing drive
Содержание Axio Scope.A1
Страница 1: ...Operating Manual Axio Scope A1 Microscope for Routine and Entry Level Research...
Страница 22: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 22 M60 2 0007 e 05 08 2 2 Systems Overview...
Страница 23: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 23...
Страница 24: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 24 M60 2 0007 e 05 08...
Страница 25: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 25...
Страница 26: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 26 M60 2 0007 e 05 08...
Страница 122: ......
Страница 123: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...