STARTUP
Carl Zeiss
Mounting Optional Components
Axio Scope.A1
64
M60-2-0007 e 05/08
3.4.10
Inserting a Modulator Disk in the
Condenser 0.9/1.25 H
•
Remove the condenser (
580H580H
Fig. 3-50/
1
) from the
condenser carrier (see chapter
581H581H
3.1.12).
If the condenser cannot be lowered sufficiently,
e.g. with mounted overview fixture, then it
might become necessary to remove it with the
stage carrier, then lower it to the stop and
remove the condenser.
•
Loosen the clamping screw (
582H582H
Fig. 3-50/
5
) of the
condenser’s dial segment (
583H583H
Fig. 3-50/
3
) with the
screwdriver (SW 3) and pull out the dial
segment.
•
Slide the modulator disk (
584H584H
Fig. 3-50/
4
) with its
two-pronged forked opening pointing forward
into the condenser. Make sure that the disk
engages in the guide on both inner sides of the
condenser. The guide serves as a stop for the
modulator disk. The pin of the disk’s clamping
screw must slide into the orientation groove of
the condenser.
•
Tighten the disk’s clamping screw with the
screwdriver (SW 3).
•
Replace the condenser into its carrier (see
chapter
585H585H
3.1.12).
Fig. 3-50
Modulator disk in condenser
0.9/1.25 H
Содержание Axio Scope.A1
Страница 1: ...Operating Manual Axio Scope A1 Microscope for Routine and Entry Level Research...
Страница 22: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 22 M60 2 0007 e 05 08 2 2 Systems Overview...
Страница 23: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 23...
Страница 24: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 24 M60 2 0007 e 05 08...
Страница 25: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 25...
Страница 26: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 26 M60 2 0007 e 05 08...
Страница 122: ......
Страница 123: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...